Diagnostic Accuracy of Ultrasonography for Diagnosing Liver Mass Taking CT Triphasic Liver as Gold Standard

Authors

  • Abeer Khalid Department of Diagnostic Radiology, Islamabad Diagnostic Center, Islamabad, Pakistan
  • Shamrez Khan Department of Diagnostic Radiology, Islamabad Diagnostic Center, Islamabad, Pakistan
  • Adeebah Mushtaq Department of Diagnostic Radiology, Islamabad Diagnostic Center, Islamabad, Pakistan
  • Urwah Khalid Mayo Hospital, Lahore, Pakistan
  • Zainab Khalid CMH, Bahawalpur, Pakistan
  • Ojala Shaheen Department of Diagnostic Radiology, Islamabad Diagnostic Center, Islamabad, Pakistan

DOI:

https://doi.org/10.70749/ijbr.v3i4.2986

Keywords:

Triphasic CT, liver, lesion, ultrasonography, malignant, benign, accuracy

Abstract

Background: Accurate distinction of benign and malignant lesions of the liver is critical for a proper clinical management. Ultrasonography is first line imaging modality that is used for hepatic lesions; however, diagnostic accuracy of ultrasound is much improved when shear wave elastography (SWE) technique is employed for the characterization of liver lesion on the basis of tissue stiffness. This study was done to ascertain the diagnostic accuracy of ultrasonography with shear wave elastography in the diagnosis of liver masses using triphasic computed tomography (CT) as the reference standard. Methods: This cross-sectional validation study carried out at Radiology Department of Islamabad Diagnostic Centre, Islamabad during 22 June 2024 to 30th December 2024. A total of 208 patients between 25 and 75 years of age with suspected focal lesions in the liver were included. All patients were scanned with ultrasonography with SWE followed by triphasic CT of the liver. Demographic data, clinical features, and findings of the imaging studies were recorded. The performance parameters of the diagnostic test such as sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and overall diagnostic accuracy were calculated using triphasic CT as a gold standard. Receiver operating characteristic (ROC) curve analysis was done to determine discriminative capacity of SWE stiffness values. Results: Triphasic CT identified malignant lesions in 152 (73.1%) patients and benign lesions in 56 (26.9%) patients. Ultrasonography with SWE correctly identified 128 true positive, 45 true negative, 11 false positive, and 24 false negative cases. The sensitivity, specificity, PPV, and NPV of ultrasonography with SWE were 84.2%, 80.4%, 92.1%, and 65.2%, respectively, with an overall diagnostic accuracy of 83.2%. ROC analysis demonstrated excellent diagnostic performance with an area under the curve (AUC) of 0.956 (95% CI: 0.931–0.980, p < 0.001). Conclusion: Ultrasonography when conducted with the use of shear wave elastography technique, reproduce high diagnostic accuracy in the discrimination of malignant and benign liver lesions. SWE provides useful quantitative information that adds to the diagnostic capability of conventional ultrasound and may prove a useful non-invasive adjunctive test in the evaluation of focal hepatic lesions.

Downloads

Download data is not yet available.

References

1. Venkatesh SK, Chandan V, Roberts LR. Liver masses: a clinical, radiologic, and pathologic perspective. Clin Gastroenterol Hepatol. 2014 Sep;12(9):1414-29.

https://doi.org/10.1016/j.cgh.2013.09.017

2. Borgheresi A, Agostini A, Pierpaoli L, Zannotti A, Capodagli-Colarizi S, Gabelloni M, et al. Diagnostic approach to focal liver lesions at cross-sectional imaging: a primer for beginners. Eur Rev Med Pharmacol Sci. 2023 Aug;27(15):7201-7225.

https://doi.org/10.26355/eurrev_202308_33293.

3. Lupsor-Platon M, Serban T, Silion AI, Tirpe GR, Tirpe A, Florea M. Performance of Ultrasound Techniques and the Potential of Artificial Intelligence in the Evaluation of Hepatocellular Carcinoma and Non-Alcoholic Fatty Liver Disease. Cancers (Basel). 2021 Feb 14;13(4):790.

https://doi.org/10.3390/cancers13040790.

4. Vernuccio F, Cannella R, Bartolotta TV, Galia M, Tang A, Brancateelli G. Advances in liver US, CT, and MRI: moving toward the future. Eur Radiol Exp. 2021; 5:52.

https://doi.org/10.1186/s41747-021-00250-0

5. Suffredini G, Gao WD, Dodd-o JM. Ultrasound Shear Wave Elastography Evaluation of the Liver and Implications for Perioperative Medicine. J Clin Med. 2024; 13(13):3633.

https://doi.org/10.3390/jcm13133633

6. Serag DM, Ebeid EA, Dessouky BA, Omer H. Role of shear wave elastography in characterization of hepatic focal lesions. Egypt J Radiol Nucl Med. 2020; 51:58.

https://doi.org/10.1186/s43055-020-00178-2

7. Gerber L, Fitting D, Srikantharajah K, Weiler N, Kyriakidou G, Bojunga J, et al. Evaluation of 2D- Shear Wave Elastography for Characterisation of Focal Liver Lesions. J Gastrointestin Liver Dis. 2017 Sep;26(3):283-290.

https://doi.org/10.15403/jgld.2014.1121.263.dsh.

8. Dhyani M, Xiang F, Li Q, Chen L, Li C, Bhan AK, et al. Ultrasound Shear Wave Elastography: Variations of Liver Fibrosis Assessment as a Function of Depth, Force and Distance from Central Axis of the Transducer with a Comparison of Different Systems. Ultrasound Med Biol. 2018 Nov;44(11):2209-2222.

https://doi.org/10.1016/j.ultrasmedbio.2018.07.003.

9. Mittal A, RR K. Evaluation of focal liver lesions using triple phase contrast computed tomography among Indian patients. Bioinformation. 2024 Oct 31;20(10):1429-1432.

https://doi.org/10.6026/9732063002001429.

10. Hafeez S, Alam MS, Sajjad Z, Khan ZA, Akhter W, Mubarak F. Triphasic computed tomography (CT) scan in focal tumoral liver lesions. J Pak Med Assoc. 2011 Jun;61(6):571-5.

11. Shahzad MA, Mushtaq S, Masood H, Masood D, Ahmed R. Accuracy of shear wave elastography in the diagnosis of hepatic focal lesions. Biol Clin Sci Res J. 2023; 2023:626.

https://doi.org/10.54112/bcsrj.v2023i1.625]

12. Akbar R, Ahmed A, Javid S. Triphasic computed tomography for characterization of liver lesions identified on ultrasonography. J Bashir Inst Health Sci. 2024;5(1).

https://doi.org/10.53576/bashir.005.01.0188

13. Hu X, Huang X, Chen H, Zhang T, Hou J, Song A, et al. Diagnostic effect of shear wave elastography imaging for differentiation of malignant liver lesions: a meta-analysis. BMC Gastroenterol. 2019 Apr 25;19(1):60.

https://doi.org/10.1186/s12876-019-0976-2.

14. Patil SD, Kamat HA, Nimbal V, Sajjan S, Muchandi R, Patil S, et al. Triple-Phase Computed Tomography and Sonoelastographic Evaluation in Focal Hepatic Lesions. J Pharm Bioallied Sci. 2024 Dec;16(Suppl 4):S3706-S3708.

https://doi.org/10.4103/jpbs.jpbs_1100_24.

15. Lee KHY, O'Malley ME, Haider MA, Hanbidge A. Triple-phase MDCT of hepatocellular carcinoma. AJR Am J Roentgenol. 2004;182(3):643-9.

https://doi.org/10.2214/ajr.182.3.1820643

16. Ferraioli G, Filice C, Castera L, Choi BI, Sporea I, Wilson SR, et al. WFUMB guidelines and recommendations for clinical use of ultrasound elastography: Part 3: liver. Ultrasound Med Biol. 2015 May;41(5):1161-79.

https://doi.org/10.1016/j.ultrasmedbio.2015.03.007.

17. Samir AE, Dhyani M, Vij A, Bhan AK, Halpern EF, Méndez-Navarro J, et al. Shear-wave elastography for the estimation of liver fibrosis in chronic liver disease: determining accuracy and ideal site for measurement. Radiology. 2015 Mar;274(3):888-96.

https://doi.org/10.1148/radiol.14140839.

18. Ozturk A, Mohammadi R, Pierce TT, Kamarthi S, Dhyani M, Grajo JR, et al. Diagnostic Accuracy of Shear Wave Elastography as a Non-invasive Biomarker of High-Risk Non-alcoholic Steatohepatitis in Patients with Non-alcoholic Fatty Liver Disease. Ultrasound Med Biol. 2020 Apr;46(4):972-980.

https://doi.org/10.1016/j.ultrasmedbio.2019.12.020.

19. Ichikawa T, Saito K, Yoshioka N, Tanimoto A, Gokan T, Takehara Y, et al. Detection and characterization of focal liver lesions: a Japanese phase III, multicenter comparison between gadoxetic acid disodium-enhanced magnetic resonance imaging and contrast-enhanced computed tomography predominantly in patients with hepatocellular carcinoma and chronic liver disease. Invest Radiol. 2010 Mar;45(3):133-41.

https://doi.org/10.1097/RLI.0b013e3181caea5b.

20. Adhikari K, Bhattarai M, Raman RK, Bhattarai S, Khadka H, et al. Triple Phase CT scan in Focal Liver Lesion with Histopathology Correlation. SSR Inst Int J Life Sci., 2024; 10(6): 6417-6421.

https://doi.org/10.21276/ssr-iijls.2024.10.6.10

Downloads

Published

2025-04-30

How to Cite

Khalid, A., Khan, S., Mushtaq, A., Khalid, U., Khalid, Z., & Shaheen, O. (2025). Diagnostic Accuracy of Ultrasonography for Diagnosing Liver Mass Taking CT Triphasic Liver as Gold Standard. Indus Journal of Bioscience Research, 3(4), 1313-1317. https://doi.org/10.70749/ijbr.v3i4.2986